RANGE OF APPLICATION

Detro® PAA is used for disinfection and cold-sterilization of surgical instruments, anesthesia and dental instruments, rigid and flexible endoscopes by immersing method. It is compatible with instruments that made from plastic, silicon, glass, acrylate, rubber.

PRODUCT PROPERTIES

  • Ready to use
  • Base and activator are mixed before using
  • High-level disinfection effect
  • 7 days solution life
  • Aldehyde and phenol free
  • Activity control by test strips

COMPOSITION
Detro® PAA base: Hydrogen peroxide, nonionic surfactant,  corrosion inhibitors, conservative and auxiliary agent Detro® PAA activator: N-acetilcaprolactam, 1-methyl-2-
pyrrolidone Active composition 0,23 peracetic acid that generated by mixing base and activator.

MICROBIOLOGICAL EFFECT

Bactericidal (incl.Tbc), fungicidal, virucidal, sporicidal in 5 minutes.

APPLICATION

  • Pour Detro PAA activator into base canister
  • Close the cover and shake thoroughly
  • After 10 minutes solution is ready to use
  • Activated solution can be used during 7 days in the case of pass pass not pass
    keeping in closed disinfection basin
  • Activated solution generates 2200 ppm peracetic acid. Solution
    can be used until 900 ppm. It need to be controlled by test strips.
  • Don’t leave instruments in solution more than effect time.

 

WARNING:

  • Don’t use aluminium and instruments that are coated with zinc
  • Keep it tightly closed package
  • In the case of contact with the eye, immediately wash plenty of
    water and consult a physician.
  • Store a dry and ventilated area and away from ignition sources